OPERATING INSTRUCTION MANUALMODEL 2200PPH ANALYZERREV. 9.1AquaMetrix Inc.22-121 Granton DriveRichmond Hill, ONCanada, L4B 3N4Tel: (800) 742-1413(905)
Model 2200P Page 83.7 Combination Probe with Preamplifier3.7.1 If the analyzer is mounted in such a position that the 10 ft. cable on thecombinatio
Model 2200P Page 94.0 Description of Functions4.1 Overview4.1.1 The Model 2200P is a microprocessor operated pH analyzer designed for industrialapp
Model 2200P Page 104.2 Calibration4.2.1 All pH systems need to be calibrated when first placed in service and thereafterfrom time to time. The frequ
Model 2200P Page 114.5 Analog Outputs4.5.1 The analog output signals consist of a non-isolated 0-1 mA, 0-5 Vdc, and isolated 4-20 mA signals. 4.5.2
Model 2200P Page 124.6.4 When in the menu mode, the display initially shows the current value of theparameters, such as the setpoint of the control re
Model 2200P Page 134.7.3 The following table describes the use of the 16 DIP switches:Bank S1Dip Switch Description ON OFF1 Selects temperature unit°
Model 2200P Page 14 4.11 Manual Override for Control Relay A4.11.1 The AUTO/OFF/ON button is used to set the operating mode of Relay A. Thisuseful
Model 2200P Page 155.0 START-UP AND OPERATION5.1 Password5.1.1 To enter the menu press MENU and the PASSWORD LED will illuminate. If the dipswitch h
Model 2200P Page 16actual value, or the buffer solution is contaminated. If you are confident in yourbuffer solution proceed by using the arrow keys t
Model 2200P Page 17 5.4 Analog Output Range Expand5.4.1 To increase the resolution of the analog outputs, the range may spread over anysection of t
TABLE OF CONTENTSSECTION DESCRIPTION PAGE1 GENERAL INFORMATION 12 SPECIFICATIONS 23 INSTALLATION 34 DESCRIPTION OF FUNCTIONS 95 START-UP AND OPERATION
Model 2200P Page 185.6 Control Relay A Relay Off Setpoint5.6.1 If you have configured the relay to control decreasing pH, the relay off setpoint wi
Model 2200P Page 195.9 Cycle Feature for Relays A and B5.10.1 The 2200P has a cycling feature for the control relays. This feature if enabled willc
Model 2200P Page 205.10 Alarm High5.11.1 The instrument is fitted with a relay, which is set to activate on both high and lowalarm conditions. The
Model 2200P Page 215.13 Temperature5.14.1 The temperature of the process can be read at any time by entering the menu andscrolling through the menu
Model 2200P Page 226.0 OPERATING HINTS6.1 Probe Care6.1.1 Keep the probe clean using the procedure recommended in the probe manual.Although the diff
Model 2200P Page 237.0 UTILITY MENU7.1 Utility Menu Functions7.1.1 The Utility Menu enables authorized personnel to perform the following:• Change
Model 2200P Page 247.3 Range Change7.3.1 The following ranges are offered for the 2200P:• 0 to 14 pH• 2 to 12 pH• 0 to 10 pH• 4 to 10 pH7.3.2
Model 2200P Page 257.4 Temperature Calibration7.4.1 The temperature channel was calibrated at the factory. It is a single pointcalibr
Model 2200P Page 267.6 Adjust 0-1 mA / 0-5 Vdc Output7.6.1 It may be desirable to fine tune the 0-1 mA / 0-5 Vdc output to take into account thecha
Model 2200P Page 278.0 DIAGNOSTICS8.1 Description8.1.1 The 2200P has diagnostic features which alerts the operator to invalid entries andmemory loss
Model 2200P Page 1MODEL 2200P pH CONTROLLERINSTRUCTION MANUAL1.0 GENERAL INFORMATIONThe AquaMetrix Model 2200P pH analyzer is a versatile industrial m
Model 2200P Page 289.0 TROUBLESHOOTING AND SERVICE9.1 Calibration / Display Problem9.1.1 If the problem is one of inability to calibrate or the di
Model 2200P Page 299.2.5 To find out whether the problem is in the sensor, or in the analyzer, use the self-testing features. Do not disconnect the s
Model 2200P Page 309.3.5 After the "ESCAPE" procedure it is necessary to do the following:a) Calibrate the temperature as described in Secti
Model 2200P Page 319.7 Parts and Accessories9.7.1 The major parts are listed below. When ordering parts please use the completepart number.Descript
Model 2200P Page 32STATEMENTS OF CONFORMITY FROM THE MANUFACTURERU.S.A.WARNING: This equipment generates, uses, and canradiate radio frequency energy
Model 2200P Page 22.0 SPECIFICATIONSDISPLAY:4 x 7 segment _” LED DisplayMEASURING RANGES:pH: 0.01 to 14.00 pHTemperature: 0°C to 100°C (32°F to 212°F)
Model 2200P Page 33.0 INSTALLATION3.1 Location3.1.1 Install the instrument within 3000 feet of where the AquaMetrix differential probe(P60 Series)
Model 2200P Page 43.4 Electrical ConnectionsThe 2200P is available in 3 different power options, 120Vac, 240Vac, and 24Vdc. Refer tofollowing sectio
Model 2200P Page 524Vdc3.4.3 To access the terminal strips open the door of the instrument and then unscrew thecaptive retaining screw near the upper
Model 2200P Page 63.5 Sensor Connections 3.5.1 Jumpers J42 are used to select which type of probe is being used. If a differentialprobe (P60) is us
Model 2200P Page 73.6 Sensor Connections; Combination Probe (P500K Series) With Temperature Compensation3.6.1 a) Both J42 jumpers located on the swi